Product Details Introduction
5L glass 50L stainless steel fermenters|bioreactors two BLBIO-5GJ -50SJ
10L 100L two stage fermenters|bioreactors BLBIO-10SJA-100SJA
50L500L two stage fermentors|bioreactors system BLBIO-50SJ-500SJ
10L50L500Lthree fermentors|bioreactors system BLBIO-10SJ-50SJ-500SJ-C
Single tank can be used independently; reliable plumbing shift can be sterilized separately; floorstanding whole rack, separate control cabinet or centralized control BLBIO-B, C-type control system is optional; can be a variety fermenter needed The combination is suitable for amplification from the lab to test flask fermentation studies of transition, but also can be used for small-scale production
Various combinations 5L-30L / 5L-50L / 10L-100L / 10L-50L / 30L-100L / 100L-1000L / 50L-5000L, etc.: Scale
Diameter height ratio: 1: 2-3; liquid filling factor: 70%
The main material is stainless steel 316L tank, the tank no dead ends; seed fermenter with a dedicated sampling pipes, discharge valve, shift. Using a large tank level perspective endoscopic observation, 12V safety depends on light, with temperature, PH, DO, foaming, liquid interface, vaccination mouth, feeding the mouth, and a number of alternate ports, all welds sturdy neat appearance .
Tank: Design pressure 0.3Mpa jacket: design pressure 0.3Mpa, optimized diversion designed to improve the uniformity of the exchange efficiency and tank temperature
Tank features: John Crane mechanical seals, intake sterile filter, exhaust condenser, including air intake system distributor, flame inoculation mouth (also doghouse and cleaning the mouth), the discharge opening, sampling ports, feeding mouth; five electrode automatic control parameters are: PH electrode, DO electrode, anti-foaming electrode, temperature probe mouth; every four cans were automatically controlled peristaltic pump up the acid / base / bubble enemy / training agents;
Motor: fermentation tank dedicated AC motor, ensure that the motor can operate in harsh environments;
Ventilation: The way deep ventilation, rotameter display, adjust the gas flow, the air tank is uniform gas distribution according to process requirements; using precision 0.2μm sterilizing filter (Dominica or Shanghai Yiming), anti- material backflow devices to ensure the safety and greatly extend the service life;
Mixing system: mechanical mixing system, SEW AC motor, Fuji inverter
Stir form: straight blade impeller (two) pitched blade impeller (a), mechanical compression type defoaming paddle (a), four-speed material plate;
Fermentation pH control: using intelligent PID control, dedicated Mettler PH electrode;
DO degree of fermentation control: using intelligent PID control, dedicated Mettler DO electrode;
Temperature control: electric heating, water cooling, circulating pump loop temperature control, air and water supply for self-escape function, over-temperature protection, with good heat exchange efficiency; intelligent control module not only ensures rapid temperature control but also realize energy .
Feeding: complete feeding system (acids, alkalis, bubble enemy, training agents), can be time quantitative feeding; four peristaltic pump can select sub-set respectively in the control interface, convenient and flexible in actual use. With a specially designed feeding pins to ensure safe operation of feeding;
BLBIO-C control system: Siemens S7-200 series PLC control system + 10 "LCD touch screen display operation, on-site direct operation (menu-driven), the entire Chinese menu and interface; sampling and analysis show browse tank data, system data read setting controls the stirring speed, temperature, medium tank, Ph value, DO value;
You can set the temperature, speed, Ph value, DO value, the upper and lower limits and have overrun alarm functions;
With ultra-fluid level alarm and automatic function, with super-bubble level alarm and automatically add defoamers function;
All alarms can be recorded incidents inspection;
Speed, temperature, Ph value, DO values, etc., and can be set to automatically shut down;
Tank medium temperature, stirring speed, Ph, DO and other parameters can be set in accordance with process requirements piecewise curve control;
Each parameter has a PID adjustment process shows, such as the set value, real value, PID set value, upper and lower limit value curve display, manual / automatic switch, online setting, etc.;
Historical data and curves of various parameters can be stored for many years, real-time curve during a power outage can re-boot of the restoration;
Friendly interface, you can switch the monitor screen dozen;
Feeding with metering function; time quantitative feeding;
System associated with a variety of control: control speed associated with dissolved oxygen, and dissolved oxygen associated with the control of feeding, feeding the ph value of the associated control and so on;
Real-time display system having a running process, data recording, data analysis
Parameters of the detection and control loops can be manually set and fixed in line on the screen, can be temperature, DO, PH sensor tuning adjustment, calibration of each peristaltic pump flow; operation have password protection, password into the operator and administrator-level password;
Automatically record fermenter boot time, shutdown time, the number of automatic run-time power, power outages, call time;
Software system with adaptive ability and self-diagnostic capabilities;
Add optional features items:
b feeding tank weighing scales weighing W d e air flow automatically controlled oxygen tank pressure bypass detection f g h online testing redox potential A (b) alcohol content online testing and control I exhaust O2, CO2 content of online testing A automatic sterilization function C tank automatic cleaning function U lid automatically upgrade function
Model Comments: BLBIO- XABC-D
X- tank total volume:
A- tank materials: -G: Glass / -S: stainless steel tank -V animal cell vaccine -C -M enzyme reactor tank
B- Transmission: -C: magnetic drive mechanical stirring / -J Direct drive mechanical agitation / -Q -S solid flow stir stir -G provide light
C - add optional features -A automatic sterilization - inaction manual manual situ sterilization sterilization -U -Y glass lid automatically upgrade
D - -A controller model for industrial control computer + Control Module - inaction B controller, for our standard controller -C 10-inch LCD touch screen + PLC -D 15-inch LCD touch screen + PLC + imports transmitter Siemens Stainless steel control box is -S